SU-FF-T-365: Practical Reconstruction Method for 3D CT-Based Brachytherapy with Shielded Colpostats
2006
Introduction and purpose.‐ The most extended worldwide HDR and PDR afterloading machines are from Nucletron, using the Plato Treatment Planning System (TPS) to perform the clinical dosimetry. The Fletcher‐Williamson (FW) vaginal colpostat applicators are widely used with these machines, they are made of a very dense material (densimet‐17) which allows shielding up to 50% of the dose. Almost all the TPS have incorporated the CT‐Based Brachytherapy (BT), in which the CT catheters and applicators reconstruction is based on contiguous CT transverse slices where the user points the catheter position. The problem with these applicators is the production of artefacts that makes the reconstruction …
Technique for routine output verification of Leipzig applicators with a well chamber
2005
The H-type Leipzig applicators are accessories of the microSelectron-HDR system (Nucletron, Veenendaal, The Netherlands) for treatment of superficial malignancies. Recently, the dose rate distributions in liquid water for the whole set of applicators using both source models available for the microSelectron-HDR afterloaders have been obtained by means of the experimentally validated Monte Carlo (MC) code GEANT4. Also an output table (cGy/hU) at 3 mm depth on the applicator central axis was provided. The output verification of these applicators by the user, prior to their clinical use, present practical problems: small detectors such as thermoluminescent dosimeters or parallel-plate ionizati…